UV/Vis Spectrophotometer with 190-1100nm DU-8600RN

  • Price:

    Negotiable

  • minimum:

  • Total supply:

  • Delivery term:

    The date of payment from buyers deliver within days

  • seat:

    Shanghai

  • Validity to:

    Long-term effective

  • Last update:

    2017-12-09 00:12

  • Browse the number:

    139

Send an inquiries

Company Profile

Guangdong Hongsheng Testing Equipment Co., Ltd.

By certification [File Integrity]

Contact:Mr. Cary Zeng(Mr.)  

Email:

Telephone:

Phone:

Area:Shanghai

Address:Shanghai

Website: http://drawell.jssytm.com/

Product details

DU-8600RN Split Beam UV/Visible Spectrophotometer

DU-8600RN spectrophotometers is split beam ratio monitoring optical system can restrain drift, Suitable for long time test. It is simple to fit a curve by using your standard with single or dual WL. It also have wavelength scanning function using PC software. It is widely used in colleges and QC labs.

* Features
Split beam ratio monitoring optical system
Low noise and Low stray light
Large LCD display (128*64 Dots), can display curve
High quality grating, detector and lamps
Data and Curve can be stored in real-time
Auto setting WL, auto Blank
Lamps can be turned on/off individually
Easy to change Pri-aligned lamps
Reinforced baseboard and bracket assures durability

Function
Photometric: T%, Abs
Quantitative: Standard Curve
System Utility
* WL Scan (Spectrum Scan) (PC)
* Time Scan (Kinetics) (PC)
* DNA/Protein Test (PC)
* Multi-WL Test (PC)
 
Specifications
 
Model DU-8600RN
Wavelength Range 190-1100nm
Spectral Bandwidth 2.0nm
Optical System Split Beam Ratio Monitoring, Blazed Holographic Grating (1200 lines/mm)
Wavelength Accuracy ±0.5nm
Wavelength Repeatability ≤0.2nm
Wavelength Setting Auto, Resolution 0.1nm
Photometric Range 0~200%T, -0.3­~3A, 0~9999C
Photometric Accuracy ±0.002A (0~0.5A), ±0.003A (0.5~1A), ±0.3%T (0~100%T)
Photometric Repeatability ≤0.001A (0~0.5A), ≤0.002A (0.5~1A), ≤0.2%T (0~100%T)
Stray Light ≤0.05%T(220/360nm)
Scan Speed High, Medium, Low. Max.2000nm/minute
Baseline Flatness ±0.0015A
Stability ±0.002A/h (500nm,0A)
Noise ≤0.2%T/3min (250/500nm,0%T); ≤0.5%T/3min (250/500nm,100%T)
Sample Compartment Accommodate 5-100mm Pathlength Cuvette
Detector Silicon Photodiode
Lamps Tungsten Lamp & Deuterium Lamp (Pre-aligned)
Display Graphic LCD (128x64 Dots)
Keypad 30-key Numeric Membrance Keypad
Output Port USB Port & RS232 Port
Printer Mini Serial Printer; PC Printer
PC Software Optional PC Scanning Software
Power Requirements AC 90-250V,50/60Hz
Dimension 545x468x245mm
Net Weight / Gross Weight 16kgs / 20kgs

Detailed Images


Related products





Certification & Plug